DeWalt Calibration Procedure for DW077
For use at locations using an On-Trak Calibration Range
DW077 LASER UNIT
Note: The following procedure can be used on a customer’s complete unit, or it can be used with the
customer’s compensator installed in the special cut-away service housing. All photos and figures show the
cutaway service housing.
A. CALIBRATION RANGE SETUP:
1. Set up the special DW077 Calibration Tripod beside the DW073 calibration stand. This calibration
procedure will not use the DW073 calibration stand.
2. Use the Tripod’s built-in bubble level to get the Tripod head approximately level.
3. The DW077 Calibration Tripod should be approximately 6m (20 feet) from the farthest On-Trak
sensor.
4. Mount the DW077 laser in a horizontal position on the tripod, turn the laser unit on and allow it to
level itself.
5. Adjust the tripod height so the laser beam strikes the farthest On-Trak sensor approximately in the
center of its detection area.
6. Boot up the same calibration software (“Calstation”) and On-Trak sensors that are used to
calibrate the DW073.
B. SOFTWARE CHECK:
Note: The laser beam must be rotating for the On-Trak sensors to work.
1. On the computer screen, click on the green “arrow button to start measuring the beam location.
2. Close the “Station Calibration Screen” window. This procedure only uses the “Cone Error
Adjustment Screen” window.
“Scan” “Spin” “Up” and “Down”
LASER KEYPAD
Vue de la page 0
1 2 3 4 5 6

Résumé du contenu

Page 1

1DeWalt Calibration Procedure for DW077 For use at locations using an On-Trak Calibration Range DW077 LASER UNIT Note: The following procedure can

Page 2

2 3. Confirm that the software is reading the beam location. 4. Try blocking the laser from reaching the farthest target.

Page 3

3 4. Rotate the laser unit on the tripod head 180 degrees, so that the “X” marking is lined up STRAIGHT AWAY from the On-Trak targets. 5. Once t

Page 4

4D. RECALIBRATING THE X-AXIS: 1. To calculate the desired target value for “Delta Position”, take the average of the two values previously written

Page 5

5RECALIBRATED. THE COMPENSATOR WILL NEED TO BE REPLACED. 9. Allow the laser unit to stabilize for about 30 seconds. 10. Using the LASER UNIT KEYP